China SK101 Automatic Opening Flash Point Meter With Atmospheric Pressure Correction Function for sale
China SK101 Automatic Opening Flash Point Meter With Atmospheric Pressure Correction Function for sale

SK101 Automatic Opening Flash Point Meter With Atmospheric Pressure Correction Function

SK101 Automatic opening flash point meter with atmospheric pressure correction function and automatic ignition The instrument adopts a large screen LCD, and the computer controls the whole process of the test. The instrument has automatic temperature rise control, automatic ignition and scanning, automatic detection of flash point locking results, printing of results, and automatic shutdown of gas source. The instrument has the advantages of accurate measurement, good

Ignition mode automatic ignition
Place of Origin China
Air source no air source is required
Packaging Details wooden case
Power supply AC 220V ± 22V 50Hz ± 2.5Hz
Cooling mode forced air cooling
Product Description

SK101 Automatic opening flash point meter with atmospheric pressure correction function and automatic ignition 

The instrument adopts a large screen LCD, and the computer controls the whole process of the test. The instrument has automatic temperature rise control, automatic ignition and scanning, automatic detection of flash point locking results, printing of results, and automatic shutdown of gas source. The instrument has the advantages of accurate measurement, good repeatability, stable performance and simple operation. It is widely used in power, petroleum, chemical, commodity inspection, scientific research and other departments

Automatic open flash point tester for gear oil hydraulic oil turbine oil ASTMD 92 Cleveland open cup method

SK101 automatic open flash point tester This open flash point automatic tester is designed and manufactured in accordance with the provisions of the national standards GB267-88 and GB3536-83 open flash point determination method.

Technical indicators

Display large screen full Chinese LCD module display.
Measurement range room temperature~400 ℃
Heating speed

in line with ASTM D92 standard;

About Heating rate from 5 to 17 C per min

Accuracy 0.1 ℃
Repeatability flash point ≤ 150 ℃, error ± 2 ℃
Flash point ≥ 150 ℃ Error ± 4 ℃
Power supply AC 220V ± 22V 50Hz ± 2.5Hz
Power 100VA
Operating ambient temperature 10 ℃~35 ℃
Operating ambient humidity ≤ 85%
Air source no air source is required
Ignition mode automatic ignition
Cooling mode forced air cooling
Overall dimension:400×260×280mm Weight:15kg

 

 

Recommended Products

Interested in this product?
Contact us now for more details and a custom quote.