Double column automatic atmospheric distillation tester
Double Column Automatic Atmospheric Distillation Tester The BD-D86-003ZII Double Column Automatic Atmospheric Distillation Tester is a sophisticated laboratory instrument designed for precise distillation analysis in industrial and manufacturing applications. Applicable Standards GB/T6536, ASTM D86, GB/T255, GB/T7534, GB/T3146 Performance Characteristics Independent operation: Each column can conduct experiments separately Intelligent heating management system ensures
Place of Origin
China
Water bath constant temperature range
0~60℃ (imported compressor refrigeration)
Packaging Details
Plywood packing
Volume detection range
0~150ml, Resolution: 0.1ml
Model Number
BD-D86-003ZII
Supply Ability
500 sets/month
Product Description
Double Column Automatic Atmospheric Distillation Tester
The BD-D86-003ZII Double Column Automatic Atmospheric Distillation Tester is a sophisticated laboratory instrument designed for precise distillation analysis in industrial and manufacturing applications.
Applicable Standards
GB/T6536, ASTM D86, GB/T255, GB/T7534, GB/T3146
Performance Characteristics
- Independent operation: Each column can conduct experiments separately
- Intelligent heating management system ensures distillation rate compliance with experimental method requirements
- User-configurable recording points for customized data collection
User-Settable Recording Points
- Record recovery volume corresponding to specific temperatures
- Record temperature corresponding to specific recovery volumes
- Automatic recording of standard national specification points
Four Experiment End Methods
- End Point: Experiment concludes when endpoint is detected
- Dry Point End: Experiment concludes when dry point is detected
- Temperature End: Experiment concludes at user-set temperature with printed output
- Volume End: Experiment concludes at user-set volume with printed output
The system features automatic atmospheric pressure detection and includes over-temperature protection for safe operation.
Technical Specifications
| Parameter | Specification |
|---|---|
| Temperature Measurement Range | Room Temperature ~ 400℃ (Resolution: 0.1℃) |
| Water Bath Constant Temperature Range | 0~60℃ (Imported Compressor Refrigeration) |
| Measuring Cylinder Peripheral Temperature | 5~50℃ |
| Distillation Rate | 4~5ml/min |
| Volume Detection Range | 0~150ml (Resolution: 0.1ml) |
| Temperature Measuring Element | PT100 (Original German JUMO Sensor) |
| Display | 1024*600 Resolution True Color Screen with Chinese Characters |
| Heating Method | Low Voltage DC Heating System |
| Printing | Embedded Micro Printer |
| Power Consumption | Less than 1kW |
| Refrigeration Method | Compressor Refrigeration |
| Ambient Temperature | 15~25℃ |
| Operation Method | Program Start with Simple Operation |
| System Management | Laboratory LIMS System Support |