Product Description Micro Surgical Instruments Surgery Set Small Bone Multi-Function Drill Saw for Veterinary veterinary saw Our electric bone surgery power system, mainly used in orthopedic trauma surgery, artificial joint replacement surgery and hand and foot surgery, neurosurgery and other fields. The product has obtained the second-class medical device registration certificate issued by the State Food and Drug Administration and the CE ,ISO13485 certification. Notes: 1.

Input voltage 110-240V 50-60HZ
Place of Origin Jiangsu, China
Warranty 2 years
Power Source Electric
Packaging Details WOODEN CASE
Quality Certification ce
Product Description
Product Description
Micro Surgical Instruments Surgery Set Small Bone Multi-Function Drill Saw for Veterinary veterinary saw
Our electric bone surgery power system, mainly used in orthopedic trauma surgery, artificial joint replacement surgery and hand and foot surgery, neurosurgery and other fields. The product has obtained the second-class medical device registration certificate issued by the State Food and Drug Administration and the CE ,ISO13485 certification.

Notes:
1. Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ),if the temperature is over 135 °C, the motor will be damaged.
2. Normal bone drill can only hold drill bits less than 6mm,acetabulum reaming drill with big torque,rpm is around 300,canulate drill is used with nail or K wire ,rpm is around 600.
3. The battery need to be get out of the power tool after surgery.Otherwise the small automatic discharge of the electric will make the life of the battery short.
4. The battery cannot be sterilized ,it is not autoclavable.
5. The cranial bur must be washed right after the surgery by purger in order to avoid the blood stick in the bur,then we should make it dry.
6. The battery space in the handpiece of the tool should not have any water,after sterilization.
7. The cranial bur blade is made of medical stainless steel,in order to avoid broken,so it is flexible and not very hard,it is
easy to be attrite,we suggest each blade only use for 1 surgery.
Details Images
Configuration
Accessory list
Quantity
Note
Handpiece
1 PC
Standard
Battery
2 PCS
Standard
Charger
1 PC
Standard
Drill chuck
1 PC
Standard
Oscillating Saw
1 PC
Standard
Blade
2 PCS
Standard
Sterilizing channel
1 PC
Standard
Aluminum alloy case
1 PC
Optional
Specification
No
Item
Value
No
Item
Value
1
Speed
0-1200rpm
6
Radial circle run-out
≤0.1mm
2
Torque
2.9N.m
7
Axial movement
≤0.5mm
3
No-load Noise
≤45dB
8
Chuck gripper diameter
≤4 mm
4
Operation(V) voltage
7.2V
9
Oscillations
0-16000 cpm
5
Sterilization temperature
155℃
10
Battery capacity
1400mAh
Company Profile