ASTM D3638 IEC 60112 Tracking Index Test Equipment 0-600V for Solid Insulating Materials
Complies with IEC 60112, UL 746A, and ASTM D3638. Features adjustable 0-600V test voltage, platinum and brass electrodes, and precise 30s drop interval. Ideal for CTI and PTI evaluation of solid insulating materials in B2B industrial applications.
ASTM Tracking Index IEC Test Equipment For Solid Insulating Materials
ASTM D3638-92 Tracking Index Test Equipment
Product information:
We have designed a series test equipment of proof and comparative tracking indices for solid insulating materials, which are strickly meet the standard requirements of IEC60112,IEC60335-1,IEC60598-1,UL746A and ASTM D3638-92 and etc..
Model specification and technical parameters :
|
Parameter model |
LT-1(0.1m³) |
LT-2(0.5m³) |
LT-3(0.5m³) |
|
Control |
Meters |
Meters |
PLC |
|
Display |
Digital |
Digital |
7" color touch screen |
|
Working voltage |
220V/50Hz or 110V/60HZ |
220V/50Hz or 110V/60HZ |
220V/50Hz or 110V/60HZ |
|
Testing voltage |
0~600V adjustable,precision 1.5% |
0~600V adjustable,precision 1.5% |
0~600V adjustable,precision 1.5% |
|
Timing device |
9999X0.1S |
9999X0.1S |
9999X0.1S |
|
Electrode |
Platinum electrode and brass electrode each for a pair |
Platinum electrode and brass electrode each for a pair |
Platinum electrode and brass electrode each for a pair |
|
Electrode dimension |
(5±0.1)×(2±0.1)×(≥12)㎜, 30°slant |
(5±0.1)×(2±0.1)×(≥12), 30°slant |
(5±0.1)×(2±0.1)×(≥12), 30°slant |
|
Electrode included angle and distance |
60°±5°,distance is (4±0.1㎜) |
60°±5°, distance is (4±0.1㎜) |
60°±5°, distance is (4±0.1㎜) |
|
Electrode pressure |
1.00N±0.05N(digital display is optional) |
1.00N±0.05N(digital display is optional) |
1.00N±0.05N(digital display is optional) |
|
The interval of dropping liquid |
30±5S,adjustable |
30±5S,adjustable |
30±5S,adjustable |
|
Dropping liquid height |
35±5mm |
35±5mm |
35±5mm |
|
Dropping liquid delay |
0.2S(solenoid valve opening time) |
0.2S(solenoid valve opening time) |
0.2S(solenoid valve opening time) |
|
Dropping liquid size |
45~50 drip/cm3 |
45~50 drip/cm3 |
45~50 drip/cm3 |
|
Time-delay circuit |
2±0.1S(in 0.5A or larger current) |
2±0.1S(in 0.5A or larger current) |
2±0.1S(in 0.5A or larger current) |
|
Short-circuit pressure drop |
8% MAX |
8% MAX |
8% MAX |
|
Short-circuit current |
1±0.1A 1% |
1±0.1A 1% |
1±0.1A 1% |
|
Wind speed |
0.2m/s |
0.2m/s |
0.2m/s |
|
Environmental requirement |
0~40℃,relative humidity≤80%,in the place of no obvious vibration and corrosive gas |
||
|
conforms to standard |
IEC60695, IEC60112,IEC60335-1,IEC60598-1,UL746A and ASTM D3638-92 |
||
Working principles :
The leakage tracking test(tracking index testing) is that the conducting liquid (0.1%NH 4 CL) of required volume in the required height (35mm ) and required time (30s) drops with the voltage between the platinum electrodes(2mm × 5mm )on the surface of solid insulating material.
Determination:
Thus the users evaluate the tracking resistance performance of solid insulating material surface under the combined influence of electric field and humid or contaminate medium. In a word, this device is used to measure the comparative tracking index (CTI) and the proof tracking index (PTI).